Cookies
We respect your privacy and are committed to protecting your personal information.
Here's how we handle your data: No tracking cookies. We do not use cookies or similar technologies to track your ...
Privacy policy
Triple Quad MS with AEI Source, SmartTune, and NeverVent Technology.
Specifications
Triple Quad MS with AEI Source, SmartTune, and NeverVent Technology.
Status: Active
Country: USA
Available configurations: Large Turbo VPI with EI-CI ExtractaBrite/Large Turbo, NeverVent with AEI/Large Turbo, No VPI with AEI/Medium Turbo, No VPI with ExtractaBrite
Analyzer type: Tandem Quadrupole MS
Scan modes: Full Scan (FS), SIM, FS/SIM (t-SIM), and FS/t-SIM; Multiple selected reaction monitoring (MRM/SRM), optional timed acquisition (t-SRM), combined SRM/FS, combined (optional) t-SRM/FS, product ion scan, precursor ion scan, neutral loss scan
Mass range: 1.2 to 1100 amu
Mass resolution/ Mass resolving power: Unit mass
Ionization Modes: Electron (EI)
Ionization Modes: Chemical Ionization (CI), Positive/Negative Chemical Ionization (PCI/NCI)
Dynamic range: >7 orders (0–110 μA)
Sensitivity: IDL ≤0.3 fg OFN
Mass accuracy/Stability: ±0.1 amu/48 h
Detector type: XLXR Detection System
Electron energy: 0–150 eV
Scanning speed: Up to 20,000 u/sec
Direct inlet option: Yes
Ion source temperature: Up to 350°C
Dwell time: 0.5 ms
MRM transition rate: Up to 800 SRM/sec
Collision energy: 0-60 eV
Installation specifications: AEI SRM: 1 µL of 1 fg/µL OFN for m/z 272>222 S/N ≥ 300:1;
ExtractaBrite EI SRM: 1 µL of 100 fg/µL OFN for m/z 272>222 S/N ≥ 16,500:1 (4,000:1);
PCI SRM: 1 µL of 5 pg/µL BZP for m/z 183>105 S/N ≥ 2,500:1 (800:1);
NCI Full Scan: 1 µL of 1 pg/µL OFN for m/z 272 S/N ≥ 10,000:1 (3,300:1)