Specifications

  • Category: Gas Chromatography - Mass Spectrometry
  • Technique: Triple Quad GC-MS
  • Manufacturer: Shimadzu
  • Country: Japan
  • Available: Discontinued
All specifications

About

Short description

Retired friple Quad GC MS/MS system with UFsweeper Collision Cell and Smart Operation Technology.

Specifications

Retired friple Quad GC MS/MS system with UFsweeper Collision Cell and Smart Operation Technology.

  • Status: Discontinued
  • Country: Japan
  • Available configurations: EI; EI/CI
  • Analyzer type : Quadrupole MS (single)
  • Scan modes : SIM, MRM, SCAN/MRM
  • Mass range : m/z 10 to 1090
  • Mass resolution/ Mass resolving power : 0.5 to 3.0 amu FWHM
  • Ionization Modes : Electron (EI)
  • Ionization Modes : Chemical Ionization (CI), Positive/Negative Chemical Ionization (PCI/NCI)
  • Sensitivity : 1 pg OFN S/N ≥600; 100 fg OFN MRM S/N ≥3000
  • Mass accuracy/Stability : ±0.1 amu/48 h
  • Detector type : Secondary Electron Multiplier with Overdrive Lens
  • Electron energy : 10–200 eV
  • Scanning speed : 20,000 u/sec
  • Direct inlet option : Yes
  • Ion source temperature : 140 to 300°C
  • Maximum column Flow Rate : Up to 10 mL/min
  • Dwell time : <1 ms
  • MRM transition rate : 16 transitions/event, 512 events, >600 MRMs/sec
  • Collision cell : UFsweeper Collision Cell
  • Collision energy : Up to 60 eV
  • Installation specifications : IDL ≤12 fg OFN S/N ≥600 (MRM)
  • Vacuum system : Turbo pump 179/185 L/s
  • Power Requirements and Consumption : GC: 1800 VA, MS: 1000 VA
  • Dimensions and Weight: 100 kg (GCMS), 10 kg (Pump)