Specifications

  • Category: Liquid Chromatography - Mass Spectrometry
  • Technique: Ion Trap/Orbitrap LC-MS
  • Manufacturer: Thermo Fisher Scientific
  • Country: USA
  • Available: Discontinued
All specifications

About

Short description

Retired LC-MS/MS system. Quadrupole-Orbitrap hybrid. High-resolution accurate-mass.

Specifications

Retired LC-MS/MS system. Quadrupole-Orbitrap hybrid. High-resolution accurate-mass.

  • Status: Discontinued
  • Country: USA
  • Vacuum system : Diff. pumped; <1×10⁻⁹ mbar; 2× turbo + rotary vane
  • Power Requirements and Consumption : 2× 230 VAC ±10% single phase, 15 A, 50/60 Hz; ~2500 W
  • Environmental requirements : 15-26°C, 40-70% RH, no condensation
  • Dimensions and Weight: MS only: 95.0 × 83.0 × 91.0 cm; 182.0 kg
  • Analyzer type : Tandem Quadrupole/Ion trap MS
  • Scan modes : Full MS, SIM, AIF, in-source fragmentation, positive/negative switching
  • Mass range : m/z 50 to 6,000
  • High resolution : Yes
  • Mass resolution/ Mass resolving power : 140,000 at m/z 200, optional 280,000
  • Dynamic range : > 5,000
  • Sensitivity : Full MS: 500 fg buspirone S/N 100:1; SIM: 30 fg S/N 100:1
  • Mass accuracy/Stability : Internal: <1 ppm RMS, External: <3 ppm RMS
  • Scanning speed : Up to 12 Hz at res 17,500
  • Dwell time : -
  • MRM transition rate : -
  • Polarity switching : One full cycle < 1 s (pos/neg)
  • Fragmentation : HCD
  • Ion source types (LC-MS) : ESI, APCI, APCI/APPI