Specifications

  • Category: Liquid Chromatography - Mass Spectrometry
  • Technique: Ion Trap/Orbitrap LC-MS
  • Manufacturer: Thermo Fisher Scientific
  • Country: USA
  • Available: Active
All specifications

About

Short description

Versatile Q-Orbitrap for broad applications, including biopharma; supports intact-protein and peptide analysis with internal calibration.

Specifications

Versatile Q-Orbitrap for broad applications, including biopharma; supports intact-protein and peptide analysis with internal calibration.

  • Status: Active
  • Country: USA
  • Vacuum system : Six-stage turbomolecular pump
  • Power Requirements and Consumption : 2×208-240 Vac, 15 A; ~3,440 W
  • Environmental requirements : 18-27 °C, 20-80% RH
  • Dimensions and Weight: 53.4 × 76.3 × 70.3 cm, 120 kg
  • Analyzer type : Quadrupole/Orbitrap
  • Scan modes : Full MS, AIF, tSIM, DIA, tMS2, ddMS2
  • Mass range : m/z 40 to 6,000 (with options)
  • High resolution : Yes
  • Mass resolution/ Mass resolving power : 240,000 FWHM at m/z 200
  • Dynamic range : > 5,000
  • Sensitivity : Full Scan: <1.5 fg reserpine; MS/MS: 100 fg S/N 100:1
  • Mass accuracy/Stability : <1-3 ppm depending on calibration type
  • Scanning speed : Up to 40 Hz at resolution 7,500
  • Dwell time : -
  • MRM transition rate : -
  • Polarity switching : < 600-700 ms per cycle
  • Fragmentation : HCD
  • Ion source types (LC-MS) : OptaMax NG (HESI), APCI/APPI optional