Specifications

  • Category: Liquid Chromatography - Mass Spectrometry
  • Technique: Ion Trap/Orbitrap LC-MS
  • Manufacturer: Thermo Fisher Scientific
  • Country: USA
  • Available: Active
All specifications

About

Short description

Linear ion trap MSⁿ. Affordable. General-purpose use.

Specifications

Linear ion trap MSⁿ. Affordable. General-purpose use.

  • Status: Active
  • Country: USA
  • Vacuum system : Differentially-pumped, split-flow turbo, dual rotary pumps
  • Power Requirements and Consumption : 230 VAC ±10%, 15 A, 50/60 Hz; ~2300 W
  • Environmental requirements : 15-27 °C; 40-80% RH
  • Dimensions and Weight: 56.0 × 79.0 × 59.0 cm; 120.0 kg
  • Analyzer type : Ion Trap MS
  • Scan modes : Full-scan, SRM, MSⁿ, ZoomScan, Ultra ZoomScan, TurboScan
  • Mass range : m/z 15 to 200; 50 to 2,000; 200 to 4,000
  • High resolution : No
  • Mass resolution/ Mass resolving power : 0.15, 0.3, 0.45, 0.7, 3 FWHM
  • Dynamic range : 4-5 orders of magnitude
  • Sensitivity : HESI/APCI MS/MS: 250 fg Reserpine ≥100:1; HESI MSⁿ: 250 fg Reserpine ≥25:1
  • Mass accuracy/Stability : -
  • Scanning speed : Turbo: 125,000 Da/s; Normal: 16,667 Da/s; Enhanced: 5,000 Da/s; Zoom: 1,111 Da/s; Ultrazoom: 28 Da/s
  • Dwell time : -
  • MRM transition rate : -
  • Polarity switching : 100 ms
  • Fragmentation : CID, PQD
  • Ion source types (LC-MS) : ESI, HESI, APCI, APPI