Specifications

  • Category: Liquid Chromatography - Mass Spectrometry
  • Technique: Ion Trap/Orbitrap LC-MS
  • Manufacturer: Thermo Fisher Scientific
  • Country: USA
  • Available: Active
All specifications

About

Short description

Hybrid quadrupole-Orbitrap MS; High-field Orbitrap analyzer; Multiple fragmentation modes

Specifications

Hybrid quadrupole-Orbitrap MS; High-field Orbitrap analyzer; Multiple fragmentation modes

  • Status: Active
  • Country: USA
  • Vacuum system : Six-stage with single turbomolecular pump + dry pump
  • Power Requirements and Consumption : 208-240 V, 15 A (instrument); 200-230 V, 15 A (dry pump); 100-240 V, 15 A (data system); ~2000 W heat output
  • Environmental requirements : 18-27 °C; 20-80% RH; ≤3,000 m altitude
  • Dimensions and Weight: 53.5 × 75.8 × 70.3 cm; 110.0 kg; without pump and PC
  • Analyzer type : Quadrupole/Orbitrap
  • Scan modes : Full MS, SIM, MS/MS, tMS², DIA, AIF, tSIM, ddMS², ddSIM, eDR scans, Hybrid-DIA
  • Mass range : m/z 40 to 6,000 (standard), up to m/z 12,000 (BioPharma Edition)
  • High resolution : Yes
  • Mass resolution/ Mass resolving power : Up to 480,000 FWHM at m/z 200
  • Dynamic range : 5 orders of magnitude (eDR mode)
  • Sensitivity : Full scan: <1.5 fg reserpine; MS²: <1.5 fg; SIM: <0.8 fg
  • Mass accuracy/Stability : <3 ppm RMS drift over 4 weeks (1-point calibration); <1 ppm RMS over 24 h (lock mass, EASY-IC)
  • Scanning speed : Up to 70 Hz (Orbitrap)
  • Dwell time : -
  • MRM transition rate : -
  • Polarity switching : < 700 ms (full scan), < 600 ms (tSIM)
  • Fragmentation : HCD, ETD, EThcD
  • Ion source types (LC-MS) : OptaMax Plus ESI, APCI, APPI; EASY-IC; Easy-Spray; NanoSpray Flex NG; FAIMS Pro Duo

Articles

Other company product

253 Plus 10kV

  • Category: Inorganic and Isotopic Mass Spectrometry
  • Technique: Isotope Ratio Mass Spectrometry for Light Stable Isotopes
  • Manufacturer: Thermo Fisher Scientific
  • Country: Germany
  • Available: Active
Detailed