Specifications

  • Category: Liquid Chromatography - Mass Spectrometry
  • Technique: Ion Trap/Orbitrap LC-MS
  • Manufacturer: Thermo Fisher Scientific
  • Country: USA
  • Available: Discontinued
All specifications

About

Short description

Retired uadrupole-Orbitrap hybrid; Routine qualitative and quantitative LC-MS/MS; Fast polarity switching

Specifications

Retired uadrupole-Orbitrap hybrid; Routine qualitative and quantitative LC-MS/MS; Fast polarity switching

  • Status: Discontinued
  • Country: USA
  • Vacuum system : Two split-flow turbomolecular pumps + one rotary vane pump; Seven vacuum regions
  • Power Requirements and Consumption : 2 × 230 VAC ±10%, 15 A, 50/60 Hz (instrument); ~2500 W avg
  • Environmental requirements : 15-26 °C; 40-70 % RH, non-condensing; ~9000 BTU/h
  • Dimensions and Weight: 95.0 × 83.0 × 91.0 cm; 182.0 kg (without pump)
  • Analyzer type : Quadrupole/Orbitrap
  • Scan modes : Full MS, SIM, PRM, DIA, AIF, In-source fragmentation
  • Mass range : m/z 50 to 3,000
  • High resolution : Yes
  • Mass resolution/ Mass resolving power : 70,000 FWHM at m/z 200
  • Dynamic range : > 5,000
  • Sensitivity : Full MS: 500 fg buspirone S/N 100:1; SIM: 50 fg buspirone S/N 100:1
  • Mass accuracy/Stability : Internal: <1 ppm RMS; External: <3 ppm RMS
  • Scanning speed : Up to 12 Hz at 17,500 resolution
  • Dwell time : -
  • MRM transition rate : -
  • Polarity switching : One full cycle (< 1 sec at 35,000 resolution)
  • Fragmentation : HCD
  • Ion source types (LC-MS) : HESI II; APCI optional; APPI optional