Specifications

  • Category: Liquid Chromatography - Mass Spectrometry
  • Technique: Ion Trap/Orbitrap LC-MS
  • Manufacturer: Thermo Fisher Scientific
  • Country: USA
  • Available: Discontinued
All specifications

About

Short description

Retired Linear ion trap. High sensitivity. Fast scanning.

Specifications

Retired Linear ion trap. High sensitivity. Fast scanning.

  • Status: Discontinued
  • Country: USA
  • Vacuum system : Diff. pumped to 10⁻⁵ Torr; turbo + rotary pumps
  • Power Requirements and Consumption : 230 VAC ±10%, 15 A, 50/60 Hz; 120/230 VAC for data system; ~2300 W
  • Environmental requirements : 15-27°C, 4-80% RH, no condensation
  • Dimensions and Weight: MS only: 56.0 × 79.0 × 59.0 cm; 120.0 kg
  • Analyzer type : Ion Trap MS
  • Scan modes : Full scan, SIM, full scan MS/MS, MSⁿ/CRM, MSn, ZoomScan, Ultra ZoomScan, TurboScan
  • Mass range : m/z 50 to 2,000
  • High resolution : No
  • Mass resolution/ Mass resolving power : Unit mass resolution
  • Dynamic range : -
  • Sensitivity : ESI/APCI: 250 fg reserpine, S/N ≥50:1
  • Mass accuracy/Stability : -
  • Scanning speed : -
  • Dwell time : -
  • MRM transition rate : -
  • Polarity switching : -
  • Fragmentation : CID
  • Ion source types (LC-MS) : ESI, APCI, APCI/APPI, NanoSpray, AP MALDI

Articles

Other company product

Trace GC Ultra

  • Category: Chromatography
  • Technique: Gas Chromatography
  • Manufacturer: Thermo Fisher Scientific
  • Country: USA
  • Available: Discontinued
Detailed