Specifications

  • Category: Liquid Chromatography - Mass Spectrometry
  • Technique: Ion Trap/Orbitrap LC-MS
  • Manufacturer: Thermo Fisher Scientific
  • Country: USA
  • Available: Discontinued
All specifications

About

Short description

Retired dual-cell linear ion trap MS; Multi-stage fragmentation; High sensitivity

Specifications

Retired dual-cell linear ion trap MS; Multi-stage fragmentation; High sensitivity

  • Status: Discontinued
  • Country: USA
  • Vacuum system : Differential pumping to 10⁻⁵ Torr; Split-flow turbo; Dual rotary pumps
  • Power Requirements and Consumption : 230V ±10%, 15A; 120/230V for data system
  • Environmental requirements : 15-27°C; RH 40-80%; Heat output 2,300 W
  • Dimensions and Weight: MS: 56.0 × 79.0 × 59.0 cm; 120.0 kg; Pumps 38.6 kg each
  • Analyzer type : Ion Trap MS
  • Scan modes : Full-scan MS; Product ion MS/MS; SRM; MSⁿ (n=1-10); ZoomScan; Ultra ZoomScan; Rapid Scan; TurboScan; WideBand Activation
  • Mass range : m/z 15 to 200; 50 to 2,000; 200 to 4,000
  • High resolution : No
  • Mass resolution/ Mass resolving power : Peak width 0.075-3 Da FWHM depending on scan type
  • Dynamic range : 6 orders of magnitude
  • Sensitivity : HESI: ≥ 100:1 (100 fg reserpine, MS/MS); APCI: ≥ 100:1; MSⁿ: ≥ 100:1
  • Mass accuracy/Stability : Mass accuracy table by scan type - Turbo 1.5 Da to Ultrazoom 0.1 Da
  • Scanning speed : TurboScan 125,000 Da/sec; Rapid 66,667; Normal 33,333; Enhanced 10,000; Zoom 2,222; Ultrazoom 28
  • Dwell time : -
  • MRM transition rate : -
  • Polarity switching : 100 ms
  • Fragmentation : CID; PQD; HCD
  • Ion source types (LC-MS) : ESI; HESI; APCI; APPI