Specifications

  • Category: Liquid Chromatography - Mass Spectrometry
  • Technique: Ion Trap/Orbitrap LC-MS
  • Manufacturer: Thermo Fisher Scientific
  • Country: USA
  • Available: Active
All specifications

About

Short description

HRAM mass detector for biopharma QC and MAM; integrates with Chromeleon and allows method transfer from Exploris 240.

Specifications

HRAM mass detector for biopharma QC and MAM; integrates with Chromeleon and allows method transfer from Exploris 240.

  • Status: Active
  • Country: USA
  • Vacuum system : Six-stage turbomolecular pump
  • Power Requirements and Consumption : 2×208-240 Vac, 15 A; ~3,440 W
  • Environmental requirements : 18-27 °C; 20-80% RH
  • Dimensions and Weight: 53.4 × 76.3 × 70.3 cm, 120 kg
  • Analyzer type : Quadrupole/Orbitrap
  • Scan modes : Full MS
  • Mass range : m/z 40 to 3,000 (to 8,000 with BioPharma option)
  • High resolution : Yes
  • Mass resolution/ Mass resolving power : 180,000 FWHM at m/z 200
  • Dynamic range : > 5,000
  • Sensitivity : Full scan <3.0 fg reserpine on column
  • Mass accuracy/Stability : <1-3 ppm depending on calibration; EASY-IC <1 ppm for ≥5 days
  • Scanning speed : Up to 22 Hz at res. 15,000
  • Dwell time : -
  • MRM transition rate : -
  • Polarity switching : < 700 ms per full cycle
  • Fragmentation : HCD
  • Ion source types (LC-MS) : OptaMax NG (HESI-II); APCI/APPI optional