Specifications

  • Category: Liquid Chromatography - Mass Spectrometry
  • Technique: Ion Trap/Orbitrap LC-MS
  • Manufacturer: Thermo Fisher Scientific
  • Country: USA
  • Available: Discontinued
All specifications

About

Short description

Retired high-field Orbitrap hybrid MS; Dual-pressure linear ion trap; Multiple fragmentation techniques

Specifications

Retired high-field Orbitrap hybrid MS; Dual-pressure linear ion trap; Multiple fragmentation techniques

  • Status: Discontinued
  • Country: USA
  • Vacuum system : Differentially pumped; multiple turbomolecular and rotary pumps
  • Power Requirements and Consumption : 230V ±10%, 3-phase, 15A; PC and chiller separate
  • Environmental requirements : 16-26°C, 50-80% RH
  • Dimensions and Weight: Without ETD: 142.0 × 87.0 × 142.0 cm; ~620.0 kg
  • Analyzer type : Orbitrap/Ion Trap MS
  • Scan modes : MSⁿ for n=1-10
  • Mass range : m/z 50 to 2,000; m/z 200 to 4,000
  • High resolution : Yes
  • Mass resolution/ Mass resolving power : 60,000 at m/z 400 (4 Hz); min 15,000; max >240,000 at m/z 400
  • Dynamic range : > 5,000 within single scan
  • Sensitivity : ESI: 100 fg reserpine, S/N ≥100:1
  • Mass accuracy/Stability : <3 ppm RMS (external); <1 ppm RMS (internal)
  • Scanning speed : 4 Hz at 60,000 resolution
  • Dwell time : -
  • MRM transition rate : -
  • Polarity switching : -
  • Fragmentation : CID, HCD, ETD (optional)
  • Ion source types (LC-MS) : H-ESI II; ESI; NanoSpray Flex; APCI; APPI

Articles

Other company product

iCAP MTX

  • Category: Inductively Coupled Plasma Mass Spectrometry
  • Technique: Triple Quad ICP-MS
  • Manufacturer: Thermo Fisher Scientific
  • Country: Germany
  • Available: Active
Detailed