Specifications

  • Category: Inorganic and Isotopic Mass Spectrometry
  • Technique: Secondary Ion Mass Spectrometry
  • Manufacturer: CAMECA
  • Country: France
  • Available: Active
All specifications

About

Short description

The Large Geometry Secondary Ion Mass Spectrometer (LG-SIMS) optimized for applications in geochronology, cosmochemistry, and environmental studies. Features include a versatile multicollector system and advanced ion imaging capabilities.

Specifications

The Large Geometry Secondary Ion Mass Spectrometer (LG-SIMS) optimized for applications in geochronology, cosmochemistry, and environmental studies. Features include a versatile multicollector system and advanced ion imaging capabilities.

  • Status: Active
  • Country: France
  • Isotopic: Yes
  • Elemental: Yes
  • Type of measurable elements: Most of elements, isotopes, and molecular species
  • Measurable isotopes/elements: Covers isotopes such as C, O, S, Li, B, Mg, and Pb
  • Analyzer type : Magnetic Sector MS
  • Ion source type: High-brightness RF plasma oxygen source and cesium source for both positive and negative secondary ions
  • Inlet types: Sample holders with automated storage chamber
  • Mass range : suitable for isotopes from Li to U
  • High Resolution : Yes
  • Mass resolution/ Mass resolving power : Up to 20,000 (50% definition)
  • Multicollector : Yes
  • Collectors qty : 7 collectors (including Faraday Cups and Electron Multipliers)
  • Dynamic range : High, allowing for simultaneous detection of isotopes with large abundance differences
  • Detection limits : ppm to ppb levels
  • Detector type : Faraday Cups (with low noise 10¹² Ω resistors optional) and discrete dynode Electron Multipliers
  • Primary beam : Cs+ and O− with high brightness for sub-µm lateral resolution
  • Spatial resolution : Sub-micron lateral resolution
  • Depth profile : High depth resolution with dynamic SIMS capabilities
  • Camera : High-resolution UV-light optical system for imaging
  • Geometry type : Double-focusing magnetic sector geometry
  • Focusing type : Double focusing with advanced ion optics
  • Vacuum system : Ultra-high vacuum (UHV) system with Pfeiffer turbo-molecular pumps
  • Gas supply: Oxygen flooding for sensitivity improvement

Articles

Other company product

ACTINIS

  • Category: Inorganic and Isotopic Mass Spectrometry
  • Technique: Secondary Ion Mass Spectrometry
  • Manufacturer: CAMECA
  • Country: France
  • Available: Active
Detailed